GE Healthcare has identified an exposed sharp edge inside the electronics cabinet of its SIGNA™ Artist and SIGNA™ Artist Evo magnetic resonance imaging systems. The protruding edge can damage power cables, potentially leading to an electrical fault.
If an electrical fault occurs and the cabinet door is opened while the system remains powered, an arc flash could be generated, posing a serious injury risk to service personnel in the immediate vicinity.
Regulatory UpdatesSep 18, 2026•3 min read
GE Healthcare alerts on exposed sharp edge in SIGNA™ Artist cabinets
GE Healthcare reports an exposed sharp edge inside SIGNA™ Artist and Artist Evo electronics cabinets that may damage power cables and cause an electrical fault or arc flash risk.
Compiled by the Pharmacountry team · Source: Health Canada