Health Canada has issued an alert for Philips Ingenia Achieva and Intera MRI systems. When exporting diffusion tensor imaging (DTI) scans using the Enhanced MR DICOM format, the exported files contain incorrect diffusion direction data. This error prevents third‑party tractography software from correctly reconstructing nerve fiber pathways in the brain.
규제 업데이트2026년 9월 17일•3분 읽기
Philips MR Systems Export Incorrect Diffusion Data in Enhanced DICOM Format
Philips Ingenia Achieva and Intera MR scanners export DTI scans with incorrect diffusion directions when using Enhanced MR DICOM, preventing proper tractography processing.
Pharmacountry 팀이 정리했습니다 · 출처: Health Canada
주제:
평점
(0)
(0)
(0)
(0)
(0)
평가 남기기